Testing Performance
To testthe single-clock. single-edge. state acquisition (logic analyzer)
3 Set up the oscilloscope. If the oscilloscope was not configured for the previous test,
then do the following steps.
a Select Setup, then select Default Setup.
b Configure the oscilloscope according to the following table.
Oscilloscope Setup
Acquisition
Averaging: On
# of averages: 16
Display
Graticule
Graphs: 2
Trigger
Level: -250 mV
[Shift] LI. Time
Stop src: channel 2 [Enter)
Channell
Alternate Scale
Attenuation: 20.00:1
Scale: 200 mVidiv
Offset -1.300 V
Channel2
Alternate Scale
Attenuation: 20.00:1
Scale: 200 mVidiv
Offset -1.300 V
Define meas
Thresholds: user-defined
Units: Volts
Upper: -980 mV
Middle: -1.30 V
Lower: -1.62 V
Set up the logic analyzer
Ana I y:er I
Uam~'
l
MACHINE I )
Type' {
StHe
J
Anlll y:er ::
Type'
(
Off
3-25