To test the voltage measurement accuracy (oscilloscope)
Testing the voltage measurement accuracy verifies the perfonnance of the following
specification:
• Voltage measurement accuracy
This test verifies the DC voltage measurement accuracy of the instrument, using a
dual cursor measurement that nullifies offset error.
Equipment Required
Equipment
DC Power Supply
Digital Multimeter
Cable
Adapter Icable to
power supply)
Adapter
Blocking Capacitor
BNC Shorting Cap
Critical Specifications
-14 Vdc to +14 Vdc, 0.1 mV resolution
Better than 0.1
%
accuracy
BNC Im)(m) 48-inch
BNC II) to Dual Banana Plug
BNC tee Im)(llIl)
BNC Im)(l) 0.18
~F,
±200 V
Recommended
ModellPart
HP 3245A option 002
HP 3458A
HP8120-1840
HP 1251-2277
HP 1250-0781
HP 10240B
HP 1250-0774
Set up the equipment
Tum on the equipment required and the logic analyzer. Let them wann up for 30
minutes
if
you have not already done so.
3-73