Testing Performance
To test the single-clock, multiple-edge, slate acquisition (logic analyzer)
3 Set up the oscilloscope. If the oscilloscope was not configured for the previous test,
then do the following steps.
a Select Setup, then select Default Setup.
b Configure the oscilloscope according to the following table.
Oscilloscope Setup
Acquisition
Averaging: On
#
of averages: 16
Display
Graticule
Graphs: 2
Trigger
Level: -250 mV
[Shift]" Time
Stop src: channel 2 [Enter]
Channell
Alternate Scale
Attenuation: 20.00:1
Scale: 200 mV/div
Offset -1.300 V
Channel 2
Alternate Scale
Attenuation: 20.00:1
Scale: 200 mVidiv
Offset -1.300 V
Define meas
Thresholds: user-defined
Units: Volts
Upper: -980 mV
Middle: -1.30 V
Lower: -1.62V
Set up the logic analyzer
1 Set up the Configuration menu.
a Press the Config key.
b In the Configuration menu, assign all pods to Machine 1. To assign all pods, select the
pod fields, then select Machine 1.
c Select the Type field in the Analyzer 1 box, then select State.
~(
Configuration
Anal y2"r ,
t1~me'
(l1ACHHl( 1 )
Type' (
Stille
)
AI' ---------------- J-
1l2' ----------------
1;-.
A:;' ---------------- L-
114'
M_
AS' ----------------
-
II';,
_
117' ----------------
rr
liS'
P_
3-50
Ana I yzer 2
Type'
(
Off
Unl,l$SI gned POdS