Set Up The Logic Analyzer - HP 1660E Series Service Manual

Logic analyzers
Table of Contents

Advertisement

CAUTION
Testing Performance
To test the offset accuracy (oscilloscope)
Set up the logic analyzer
1 Set up the Configuration menu.
a Press the Config key. At the pop up menu, select Scope Charme!.
b Select the Input field, then select Cl.
c Move the cursor to the Probe field, then use the RPG knob to dial in 1:l.
d Move the cursor to the VlDiv field, then use the PRG knob to dial in 4.00 V.
e Move the cursor to the Offset field. Set the offset to 0 by typing 0, then pressing the
Select key.
,-
f
Select the Coupling field, then select 1MQ / DC.
Set the Charmel Coupling field to 1MQ / DC or damage to the equipment will resuit.
g Move the cursor to the slDiv field, then use the RPG knob to dial in 500 ns.
~
Scope
Ch~nnel
(
AUlo~cole
) _
~
fYJ (
V/[)!v
' ) 1
01
*. .
1
l
Gl
fCOUPlin?
1 f
Preset
CI
·1.0(''/
0'1
1'1
lMs<!OC
U~er
~jD1V
1
~
t
Display
1
ffa~'PleITDoto
acquired at'
D
,
SOD
(lS
(,
s
Options
Penod
11,,1:( OCqUlsIllOIP
1 ns
"
"
2 Set up the Display menu.
a Press the Display key.
b Select the Mode field, then select Average.
c Move the cursor to the Average # field. Type 32 on the front-panel keyboard, then
press Done.
d Select the Grid field and set it to On.
e In the Waveform menu, delete charmel2.
If
charmel1
is
not inserted, insert charmell.
~ (
Scope Display
)
( I'tU\OsCale)
_
~
f
MOd",
If
Average
"lfeOnneC!
OO\Sl~
AveraQe
'2
Ofl
On
_~~:IV:l ~
f
(ilSPloYlISamp\~
j
('010 OC'lulred ot
,
,
~~' IlS
0
S
Opllons
?enod
tie::l OCqUJSl tlor"
I
ns
"
i
3-78