Table 3 Materials Required To Construct The Test Connectors; To Construct The Test Connector (For Use With Mso Models Only) - Agilent Technologies 6000 Series Service Manual

Hide thumbs Also See for 6000 Series:
Table of Contents

Advertisement

To construct the test connector (for use with MSO models only)

6000 Series Oscilloscopes Service Guide
Agilent 6000 Series Oscilloscopes that have digital channels
enabled require the test connector described below. Follow the
steps to build the test connector.
Table 3
Materials required to construct the test connectors
Description
Recommended Part
BNC (f) Connector
Agilent 1250-1032 or
Pomona 4578
Berg Strip, 8-by-2
Jumper wire
1 Obtain a BNC connector and an 8-by-2 section of Berg strip.
2 On one side of the Berg strip, solder a jumper wire to all of
the pins (shown in
3 On the other side of the Berg strip, solder another jumper
wire to all of the pins.
4 Solder the center of the BNC connector to a center pin on one
of the rows on the Berg strip.
5 Solder the ground tab of the BNC connector to a center pin
on the other row on the Berg strip.
Testing Performance
Qty
1
1
Figure 1
on page 30).
2
29