6000 Series Oscilloscopes Service Guide
3 Use the BNC tee and cables to connect the oscilloscope
calibrator /power supply to both the oscilloscope and the
multimeter (see
Oscilloscope
BNC Tee
BNC (f) to dual
banana
Figure 3
4 Adjust the output so that the multimeter reading displays the
first Volts/div calibrator setting value in
Wait a few seconds for the measurement to settle.
5 Press the Y2 softkey, then position the Y2 cursor to the center
of the voltage trace using the Entry knob.
Figure
3).
Oscilloscope
Calibrator
Connect equipment for voltage measurement accuracy test
Testing Performance
Digital
Multimeter
Table
7.
2
39