To verify basic oscilloscope operation
6000 Series Oscilloscopes Service Guide
For 6000L Series oscilloscopes, connect the oscilloscope to your
network or establish a point-to-point connection, and use the
Remote Front Panel feature described in the 6000 Series
Oscilloscopes User's Guide to control the oscilloscope. If you do
not have the user's guide, you can download it from
www.agilent.com/find/6000.
1 Press the Save/Recall key on the front panel, then press the
Default Setup softkey under the display. The oscilloscope is
now configured to its default settings.
2 Connect an oscilloscope probe from channel 1 to the Probe
Comp signal terminal on the front panel.
3 Connect the probe's ground lead to the ground terminal that
is next to the Probe Comp terminal.
4 Press AutoScale.
5 You should see a waveform on the oscilloscope's display
similar to this:
Troubleshooting
4