To Remove The Batteries (Option Bat Only); Figure 42 Removing The Batteries - Agilent Technologies 6000 Series Service Manual

Hide thumbs Also See for 6000 Series:
Table of Contents

Advertisement

To remove the batteries (Option BAT only)

6000 Series Oscilloscopes Service Guide
1 Perform the previous procedures:
"To remove the cabinet"
2 Remove the two T15 screws from the battery enclosure lid,
and open the battery enclosure lid.
3 Pull upward on the battery straps to remove the batteries.
Figure 42
Removing the batteries
Replacing 6000A Assemblies
on page 94
5
121