Figure 10 User Calibration Cable For 4-Channel Oscilloscope - Agilent Technologies 6000 Series Service Manual

Hide thumbs Also See for 6000 Series:
Table of Contents

Advertisement

To Channel 1
To Channel 2
6000 Series Oscilloscopes Service Guide
BNC(f)-to-BNC(f) (barrel connector) to the tee as shown
below.
Figure 10
User Calibration cable for 4-channel oscilloscope
3 Connect a BNC cable (40 inches maximum) from the TRIG
OUT connector on the rear panel to the BNC barrel
connector.
4 Press the Utility key, then press the Service softkey.
5 Begin the Self Cal by pressing the Start User Cal softkey.
6 When the User Cal is completed, set the rear-panel
CALIBRATION switch to PROTECTED.
Calibrating and Adjusting
To Channel 4
3
To Channel 3
Longer cable
to TRIG OUT