To remove the fan

WA R N I N G
Fan cable
6000 Series Oscilloscopes Service Guide
1 Perform the previous procedures:
"To remove the cabinet"
"To remove the power supply shield"
"To remove the power supply"
2 Remove the main shield covering the system board by sliding
it toward the back of the instrument.
Thin sheet metal parts may have sharp edges. Handle with care to avoid injury.
3 Disconnect the fan cable from the system board.
Figure 44
Removing main shield and disconnecting fan cable.
Replacing 6000A Assemblies
on page 94
on page 118
Main shield
on page 114
125
5