Advertisement

Retaining Hooks
6000 Series Oscilloscopes Service Guide
Deck
Figure 25
Removing the display assembly
7 To reinstall the display:
a Align the top locating tab with the locating slot in the
sheet metal and the retaining hooks with their retaining
holes.
b Push down on the LCD until it is flat with the deck and at
the same time push the LCD to your left.
c Ensure the latch tab is seated in its hole as shown in
Figure
25.
d Route the cables down through the deck hole and
reconnect to the system board as shown in
Replacing 6000A Assemblies
Display (LCD)
Assembly
Latch Tab
Latch Slot
Figure
5
24.