To Remove The Power Shaft; Figure 57 Removing The Power Shaft - Agilent Technologies 6000 Series Service Manual

Hide thumbs Also See for 6000 Series:
Table of Contents

Advertisement

6
Replacing 6000L Assemblies

To remove the power shaft

C A U T I O N
144
1 Perform the previous procedures:
"To remove the bottom cover"
"To remove the power supply"
"To remove the fan"
2 Use a flat-blade screwdriver to gently spread the latch while
pushing the power shaft forward.
Twisting the latch too much could cause it to break!
Figure 57
Removing the power shaft.
3 Lift and remove it from the deck.
4 When reinserting the power shaft, push the shaft into the
power switch until the shaft snaps onto the switch.
on page 133
on page 138
on page 140