Testing Performance

To construct the test connector (for use with MSO models only)

26
Agilent 7000 Series oscilloscopes that have digital channels
enabled require the test connector described below. Follow
the steps to build the test connector.
Materials required to construct the test connectors
Table 3
Description
Recommended Part
BNC (f) Connector
Agilent 1250-1032 or
Pomona 4578
Berg Strip, 8-by-2
3M .100" x .100" Pin
Strip Header or similar
Jumper wire
1 Obtain a BNC connector and an 8- by- 2 section of Berg
strip. A longer strip can be cut to length using wire
cutters.
2 On one side of the Berg strip, solder a jumper wire to all
of the pins (shown in
3 On the other side of the Berg strip, solder another jumper
wire to all of the pins.
4 Solder the center of the BNC connector to a center pin on
one of the rows on the Berg strip.
5 Solder the ground tab of the BNC connector to a center
pin on the other row on the Berg strip.
Qty
1
1 strip, cut to length (8x2)
Figure 1
on page 27).