Figure 8 Connect Equipment For External Trigger Sensitivity Test (4-Channel Models) - Agilent Technologies 6000 Series Service Manual

Hide thumbs Also See for 6000 Series:
Table of Contents

Advertisement

50Ω Feedthrough*
N to BNC adapter
6000 Series Oscilloscopes Service Guide
1 Connect the equipment (see
a Use the N cable to connect the signal generator to the
power splitter input.
b Connect one output of the power splitter to the Aux Trig
input through a 50Ω feedthrough termination.
c Connect the power sensor to the other output of the power
splitter.
Oscilloscope
Power Splitter
Power Sensor
Power Meter
Figure 8
Connect equipment for external trigger sensitivity test
(4-channel models)
Testing Performance
Figure
8).
Signal
Generator
N Cable
* Required for all models.