6
Replacing 6000L Assemblies

To remove the power supply

Power to System
Board
Power Input
Board Cable
138
1 Perform the previous procedures:
"To remove the bottom cover"
"To remove the system board"
2 Disconnect the ground cable and AC cables from the power
supply using a needle nose pliers if necessary.
3 Press the release latch to disconnect the power input board
cable from the power input board.
Power Input
Board
Figure 53
Removing the power supply.
6000 Series Oscilloscopes Service Guide
on page 133
on page 136
T10 Screws
AC
Connector
Ground
Connector
Hole in
Divider
Snap
Fasteners