5
Replacing 6000A Assemblies
System Board
104
Backlight Inverter Cable
Main Shield
Removing the main shield and disconnecting the display ca-
Figure 24
bles
4 Using a flat-blade screwdriver, lift the latch tab at the bottom
of the display assembly just enough to clear the slot in the
deck.
5 Push the entire display assembly to the right to release the
retaining hooks from their slots in the deck
6 Lift and remove the display assembly.
Display (LCD) Cable
6000 Series Oscilloscopes Service Guide