6000 Series Oscilloscopes Service Guide
Table 6
Equipment Required to Verify Voltage Measurement Accuracy
Equipment
Critical Specifications
Oscilloscope
14 mV to 35 Vdc,
Calibrator
0.1 V resolution
Digital multimeter
Better than 0.01% accuracy
Cable
BNC, Qty 2
Shorting cap
BNC
Adapter
BNC (f) to banana (m)
Adapter
BNC tee (m) (f) (f)
Blocking capacitor
1 Set up the oscilloscope.
a Adjust the channel 1 position knob to place the baseline at
0.5 major division from the bottom of the display.
b Set the Volts/Div setting to the value in the first line in
Table
7.
Testing Performance
Recommended
Model/Part
Fluke 5820A
Agilent 34401A
Agilent 10503A
Agilent 1250-0774
Agilent 1251-2277
Agilent 1250-0781
Agilent 10240B
2
37