To remove the backlight inverter board and e-field shield
6000 Series Oscilloscopes Service Guide
Some older oscilloscope models may not have the e-field shield
mentioned in this procedure.
1 Perform the previous procedures:
"To remove the cabinet"
•
"To remove the power supply shield"
•
2 Press to release the top tab and use the hole to aid in
releasing the side tab. Then open the shield. Do not bend it
past the vertical position as shown.
Figure 27
Opening the backlight inverter shield
3 Remove the LCD cables from the cable guides.
4 Release the top latch and lift the top of the board off the top
alignment post.
5 Release the bottom latch and lift the board off the bottom
alignment post.
Replacing 6000A Assemblies
on page 94
on page 114
5
107