To remove the battery controller board (Option BAT only)

6000 Series Oscilloscopes Service Guide
1 Perform the previous procedures:
"To remove the cabinet"
"To remove the batteries (Option BAT only)"
"To remove the power shaft"
2 Disconnect the DC cable from the battery controller board (to
the system board).
3 Disconnect the keyboard cable from the battery controller
board (to the front panel assembly).
4 Remove the one T15 screw securing the battery connection
board to the battery enclosure.
5 Remove the two T10 and one T15 screws securing the battery
controller board to the deck.
6 Slide the battery controller board away from the battery
enclosure to the board's slot opening, and lift the board
upward to remove it from the deck.