To Remove The Keyboard Assembly - Agilent Technologies 6000 Series Service Manual

Hide thumbs Also See for 6000 Series:
Table of Contents

Advertisement

5
Replacing 6000A Assemblies

To remove the keyboard assembly

100
1 Perform the following procedures:
"To remove the cabinet"
"To remove the front panel assembly"
2 If removing the softkey pad only skip steps 3, 4c, and 4d
below.
3 Remove all of the knobs by pulling them straight out. You may
need to use a flat-blade screwdriver to gently pry them as you
pull.
Using a twisting motion rather than prying prevents marking
or damaging the front panel.
4 Remove the main keyboard and softkey board as follows.
a Lift the left end of the softkey board enough to clear the
tab holding it in place.
b Slide the softkey board to the left to release it from the
retaining tabs.
c Release the 8 latches holding the main keypad board to the
front panel.
d You will notice the latches do not all face the same
direction. This is shown with arrows in the following
figure.
e Lift the board up just enough to clear the latches.
f Lift both boards out being careful not to damage the
ribbon cable between them.
5 Remove and replace keypads as needed.
on page 94
on page 97
6000 Series Oscilloscopes Service Guide

Advertisement

Table of Contents
loading

Table of Contents