Mitsubishi Electric MELSEC iQ-R Series User Manual page 64

Temperature control module
Hide thumbs Also See for MELSEC iQ-R Series:
Table of Contents

Advertisement

■Procedure for executing the sensor correction (one-point correction) (when using a program)
Configure the setting by the following procedure:
1.
Set 'Sensor correction function selection' (Un\G564) to Normal sensor correction (one-point correction) (0). ( Page
311 CH1 Sensor correction function selection)
2.
Set a value for 'CH1 Sensor correction value setting' (Un\G565). ( Page 312 CH1 Sensor correction value setting)
Sensor two-point correction function
With this function, an error between the temperature process value (PV) and the actual temperature between the two points
selected in advance (a correction offset value and a correction gain value) is stored. Based on this gradient, the error between
the temperature detected by a sensor and the actual temperature is corrected.
Configure the sensor two-point correction setting in the setting mode ('Setting/operation mode status' (X1): OFF). Set
"Monitor" for "Stop mode setting" of "Control basic parameters" in "Application Setting."
Temperature
process value (PV)
After
correction
(1)
(2)
(1) 'CH1 Sensor two-point correction offset value (corrected value)' (Un\G569)
(2) 'CH1 Sensor two-point correction offset value (measured value)' (Un\G568)
(3) 'CH1 Sensor two-point correction gain value (corrected value)' (Un\G571)
(4) 'CH1 Sensor two-point correction gain value (measured value)' (Un\G570)
■Procedure for executing the sensor correction (two-point correction) (when using the
engineering tool)
[Tool]  [Module Tool List]
1 FUNCTIONS
62
1.19 Sensor Correction Function
(3)
(4)
Before
correction
Input temperature
1.
Select "Temperature trace" of
"Temperature Control Module" and
click the [OK] button.

Advertisement

Table of Contents
loading

This manual is also suitable for:

R60tctrt2tt2R60tcrt4R60tctrt2tt2bwR60tcrt4bw

Table of Contents