To remove and replace the front panel bezel and front panel circuit
assembly
Disconnect the power cable and remove the cover, top plate, and bottom plate as
1
described above in the section on removing the cover, top plate, and bottom plate.
Disconnect front panel cables.
2
Grasp and pull on all 13 knobs located on the front panel to remove them (Figure 5-12).
3
Figure 5-12
Use a screw driver to disengage locking tabs located along the edges of the front panel
4
bezel (4 tabs on the side, 1 tab on the top, 1 tab on the bottom) - See Figure 5-13. Once
the tabs are disengaged, pull the front panel bezel off (Figure 5-14). Suggestion: It
tends to be easiest to disengage the side tabs and the top tab. Then pull the bezel off
at an angle and flex it until the bottom tab disengages as well.