Composite
Composite
When set to Enabled or Required, will attempt to decode the composite component of a
linear symbol. The linear symbol can be DataBar-14, DataBar Expanded, DataBar Limited,
EAN-128, UPC-A, EAN-13, EAN-8, and UPC-E.
Usage:
Allows reading by both linear and 2D readers.
Definition:
Combines 2D and linear width modulated symbology on the same symbol
where different messages can be read by each reader type.
Serial Cmd:
<K453,mode,separator status,separator>
Default:
Disabled
Options:
0 = Disabled
Enabled
If set to Enabled, the Imager only decodes the linear component, and it will only output the
linear component.
Required
If set to Required, the Imager must decode both the linear and the composite components;
otherwise it outputs a No Read.
Separator Status (Composite)
Usage:
Allows the user to distinguish between the main and Supplemental symbols.
Definition:
Separates the linear and the composite component.
Serial Cmd:
<K453,mode,separator
Default:
Disabled
Options:
0 = Disabled
Separator Character (Composite)
Usage:
As required by the application.
Definition:
Allows the user to change the separator character from a comma to a new
character.
Serial Cmd:
<K453,mode,separator status,separator>
Default:
,
(comma)
Options:
Any ASCII character.
Important: The ASCII characters <, >, and
For <:
<K453h,,,3C>
For >:
<K453h,,,3E>
,
For
:
<K453h,,,2C>
See
ASCII Character Entry Modifier
enter ASCII characters as hex values.