Attaching The Fiver Adapter And External Light Source - Nikon eclipse lv150 Instruction Manual

Industrial microscope
Table of Contents

Advertisement

2. Replacing the lamp
The lamp can be removed without having to detach the lamphouse from the microscope.
Before performing the following procedures, turn off the power supply for the microscope (press
the " " side) and unplug the power cable from the wall outlet. And check that the lamp and
lamphouse have cooled down sufficiently.
1 Loosen the lamphouse cover clamp screw using the hexagonal wrench.
2 Remove the lamphouse cover.
3 Push down the lamp clamp lever and remove the old lamp.
4 With the lamp clamp lever held down, insert the electrodes of a new lamp into the pin holes
of the socket. Press the lamp as far as it goes, and then release the lamp clamp lever to secure
the lamp.
Be careful not to touch the glass surface with bare hands.
When releasing the lamp clamp lever, use care so that the lamp does not tilt.
5 Close the lamphouse cover and secure it by tightening the clamp screw.
5

Attaching the Fiver Adapter and External Light Source

If the brightness of the specified light source is less than the desired brightness, an external light
source can be used with the LV-UEPI2 to perform the epi-fl microscopy.
The following light sources can be attached through the light guide fiber when the optional LV-
HGFA HG fiber adapter is mounted on the light source mount part.
• External light source:
54
1
2
EXFO X-Cite 120 (manual type) or EXFO X-Cite 120PC
(motorized type)