II. Microscopy
1. Microscope system consisting of TE2000-U,
dia-illuminator 100W, system condenser,
and T-TD eyepiece tube D
The following instructions are based on a microscope equipped with the parts listed below.
TE2000-U microscope
T-TD eyepiece tube D
CFI 10X eyepiece
T-DH dia-illuminator 100W
LHS-H100P-1 12V100W lamphouse
12V100W halogen lamp
TE2-PS100W power supply
System condenser
LWD condenser lens
Condenser module for bright-field microscopy
Condenser module for phase-contrast microscopy
T-SR rectangular stage
T-N6 sextuple nosepiece
Objectives for bright-field microscopy
Objectives for phase-contrast microscopy
Others
1-1. Bright-field (BF) microscopy
Key point: Detach all optical elements required for other types of observation from the optical path. Position the
condenser correctly (adjust the focus and centering), and adjust the clarity of the image by moving the
"aperture diaphragm".
1.
Reset the convenient functions.
1) Turn the "condenser refocusing clamp" on the dia-
illuminator counterclockwise to release.
2) Set the "eyepiece tube turret" to position < O >.
3) Set the "intermediate magnification dial" on the right side
of the microscope to position < 1X >.
4) Turn the "objective refocusing ring" behind the fine/coarse
focus knob on the right side of the microscope
counterclockwise to release.