7

Episcopic Illuminator

The illuminators which can be used for this microscope system include a coaxial episcopic illuminator, which is attached
between the zooming body and tube, a ring fiber illuminator, which is attached to the tip of the objective, and an illuminator
that has a flexible fiber for adjusting the illumination direction.

P2-CIA Coaxial Epi Illuminator

7.1
Attach this illuminator to the position between the zooming body
and the tube. The C-FLED2 LED Light Source for Fiber
Illuminator is used as its light source. The light from the light
source is led into the device by the two-branch fiber and episcopic
illumination is generated along the optical axis. This illuminator is
suitable for observing mirror-like samples.
 1/4 λ plate
When using coaxial epi illuminator with a 0.5x or 1x objective,
an optional 1/4 λ plate must be attached to the tip of the
objective.
1/4 λ plates are supplied to prevent flare in the imaging
system due to light reflected into the optical system.
There are the following two types of 1/4 λ plates.
• P2-CIA QL0.5X 1/4 λ plate 0.5x for coaxial epi illuminator:
Intended for 0.5x objectives
• P2-CIA QL1X 1/4 λ plate 1x for coaxial epi illuminator:
Intended for 1x objectives
When using a 0.5x objective, remove the protection ring from
the tip of the objective, and then screw the 1/4 λ plate. When
using a 1x objective, remove the correction glass from the tip
of the objective, and then screw the 1/4 λ plate.
The brightest illumination is generated when the white dot on
the 1/4 λ plate is at the front. Select a position that has the
optimum brightness and contrast for observation.
Restriction
The type of objectives used for the coaxial epi illuminator and the
recommended zoom magnification are shown below.
Objective
P2-SHR Plan Apo 0.5X
P2-SHR Plan Apo 1X
P2-SHR Plan Apo 1.6X
P2-SHR Plan Apo 2X
* Shading or vignetting may occur in the low zoom magnification range.
For details about using the light source, refer to the instruction manual for the light source.
Chapter 8
Functions and Operations of the Devices
Usable objectives and recommended zoom magnifications
SMZ18 Zooming Body
Coaxial epi
illuminator
1/4 λ plate
White dot
P2-CIA Coaxial Epi Illuminator
Zoom magnification
From 3.8x
From 4.0x
Unusable
Unusable
106
SMZ25 Zooming Body
From 2.5x
From 4.0x
Unusable
Unusable