CALF—Calibration Functions
ST: In case of the ST series, do not change the settings. If changed, weighing will not be performed properly.
Function #
CALF-01 ST
0
None
1
g
2
kg
3
t
CALF-02 ST
0
None
1
1
10
2
2
10
3
3
10
4
4
10
CALF-03 ST
Minimum division (D) for the weight. Input 1, 2, 5, 10, 20, or 50(decimal point ignored).
CALF-04 ST
Capacity of the weighing. Weighing is allowed up to this setting + 9D(9 divisions). A weight beyond
this limit is an overflow and not displayed.
CALF-05
A range of accepting "zero" from the [ZERO] key or Control I/O. Represented in terms of percent (%)
with respect to the weighing capacity, centering to the zero calibrated point in calibration. If this is set
to 2, "zero" can be accepted within a range of the zero calibrated point +/- 2 %.
Function name
Weighing unit
Decimal point position
1 2 3 4 5
1 2 3 4.5
1 2 3.4 5
1 2.3 4 5
1.2 3 4 5
Minimum division
Capacity
Zero range
1, 2, 5, 10, 20, 50 (D)
0 – 800000
Decimal point ignored
C-12
Range
0 - 30
Default
2
0
1
10000
2