4-4 Gravity compensation

Press [CALF] while pressing [ENTER] in the normal mode, and press [ENTER] next.
Set"
is displayed. (ready to calibrate)
*
Press [ESC] to return to the normal mode.
Press [FUNC.].
sub-display..
*
Press [ESC] to return to the normal mode.
3. Enter the know acceleration rate for your location
Refer to {Appendix D: Gravity acceleration map} and enter the value on the map.
Press [ENTER] to save the calibration data.
*
Press [+/-] to cancel the input and re-input.
*
Press [ESC] to return to the normal mode.
1. Get in the calibration mode
2. Get in the Gravity compensation mode
"GrAvity"
is displayed on the main display and your "g" value is on the
4-3