2-2 Wiring

Open the front panel door removing screws on the front panel, and connect a power cable, load cell
cables, and control signal cables to the terminals inside the case. Follow the instructions below.
Cable used
Load cell
Control signal cable
Power cable
Terminal
E
N
L
Load cell cable
Terminal
34
35
36
37
38
39
40
The use of a 6 wire shielded cable is recommended in order to reduce
weighing error. (When using a 4-wire cable, connect pin # 34 to # 35, and
pin # 36 to # 37.)
If the system requires two or more load cells, use an explosion-protected
type of summing box to input the signal into the indicator.
Use a shielded cable and connect its shield to pin # 40 (Frame ground).
Description
EXC+
SEN+
SEN-
EXC-
SIG+
SIG-
F.G.
Description
Ground
AC power supply
Excitation +
Sense +
Sense -
Excitation -
Signal +
Signal -
Frame ground