AND Explosion Protected Weighing Indicator AD-4403-FP ST Series Instruction Manual page 102

St series weighing indicator for explosion protected platform scale
Table of Contents

Advertisement

Function #
CALF-14
1
Normal batching (Customer Programmed Control Mode)
2
Loss-in-weight (Customer Programmed control Mode)
3
Normal batching (Built-in automatic program mode)
4
Loss-in-weigh (Built-in automatic program mode)
5
Nozzle Controlled Weighing Mode (Built-in automatic program mode)
6
Check weighing 1
7
Check weighing 2
8
Check weighing 3
9
Check weighing 4
CALF-15
When disabled, preset tare can not be set nor recalled.
0
Disable
1
Not disabled
CALF-16
When "1" is set (distinguished), " " illuminates on the lower status display
0
None (same)
section when "preset tare" is used and header of serial output becomes "PT"
(preset tare) and "T " (tare)
1
Distinguished
*
CALF-17
0
Not printed
1
Print
CALF-18
0
Both push-zero and tare-clear are disabled.
1
Only push-zero is available.
2
Only tare-clear is available.
3
Both push-zero and tare-clear are available.
Function name
Weighing mode
Preset tare
Distinguish between preset tare and tare
Pressing [PRESET TARE] always shows preset value regardless of this
setting.
Serial interface; Print preset tare value with net
weight (Si F-01=3, rsF-01=3)
Push-zero and tare-clear function when powered on
and when the display is turned on.
Push-zero is performed at stable reading. If
unstable for 10 seconds or longer,
St"
(Error message) appears on the
display.
When reading is out of the range of
push-zero,
the display.
(Press [ESC] to return to the normal mode.)
C-15
Range
See {6. Weighing}
in detail.
"Err PZr"
is displayed on
Default
3
1
0
0
0
"Err

Advertisement

Table of Contents
loading

Table of Contents