The function assigned to terminals
The function of the terminal can be assigned arbitrarily. Refer to "10.4. Parameter
List" of the function list
8.1.2.
8.1.2.
8.1.2.
8.1.2.
Timing Chart
Timing Chart
Timing Chart
Timing Chart
Caution
Keep the delay time to avoid abnormal-operation and noise.
Keep the input signal more than 40 ms to avoid noise and chattering.
Set the Communication mode in [01 f- 3] [without 5].
The transmission of BCD data synchronizes with displaying it in the following modes.
Stream mode, auto print mode, manual print mode and total print.
Exmaple
Approx. 10 ms from new
data edge to strobe signal.
Data output
Strobe signal
Hold input command
AD-4402
Approx. 20 ms width
Max. 50 ms delay
time until holding data
Page 75
Max. 50 ms delay time to
release the hold command.
10ms/div