Saving instrument settings

When placing an instrument back in the holder, the instrument settings
made are always saved in the user profile (DropMode).
The next time the instrument is removed, the last settings made are
available again.
The instrument settings speed or intensity, maximum torque if the Endo
function is activated, and activating spray (only with C+ electronic foot
control) can be saved on one of the three favorite keys. The instrument
settings can be called up again by pressing the key or the C+ foot
3
switch, see section "Setting the speed"  [→ 107], "Setting the intensity" 
[→ 138] and "Endo function"  [→ 123].
40
4.5.5.4

Setting the amount of spray water

The amount of spray air or water is factory preset for the instruments.
This basic setting and the compressed air for the turbines can be
changed by the service technician.
The amount of spray water can be adjusted for each individual
instrument by the user. The amount of air remains constant. To do this,
rotating water controls are located under the instrument holder of the
dentist element.
If the control for an instrument is completely closed, the treatment site is
cooled only with air.
1. Rotate the water knob on the respective instrument.
Reduce water quantity in spray: Turn knob clockwise to close
Increase water quantity if spray: Turn knob counterclockwise to
open
2. Test the setting for spray water. Correct the setting if necessary.
4 Operation
4.5 Dentist element
93