4 Operation
4.4 Dentist element
2
Setup
48
4.4.6

Turbine / air motor / other air-driven instruments

IMPORTANT
Please see the operating instructions for the various high-speed
handpieces, air motors, or air-driven instruments.
The high-speed handpiece hose has a standardized coupling as per
ISO 9168.
If your simulation unit is equipped with the pneumatic foot control, the
drive air for the high-speed handpiece can be regulated by the foot
pedal.
4.4.6.1

Using the highspeed handpiece

User interface SIM dentist element
When the high-speed handpiece is removed, a short vertical bar
appears on the SIM display of the dentist element. Step on the foot
control to start the instrument.
4.4.6.2

Setting the light of the highspeed handpiece

The instrument light supply is optional in Sim Intego.
Halogen lamps cannot be operated with the Sim Multi Bench simulation
unit.
The voltage or current can be limited for the LED lamp of the high-
speed handpiece. The power supply for the instrument light can also be
switched off.
The original Sirona LED lamp is usually operated at 3.6 V.
For lamps made by other manufacturers, the settings may have to be
adjusted.
NOTE
The operating voltages and operating currents of different lamps may
vary.
Overvoltages or excess current can lead to damage.
➢ When changing lamps, make sure that the operating voltage or
operating current is properly set for the new lamp.
4.4.6.3

Setting the high-speed handpiece light on the SIM dentist element

Accessing the high-sped handpiece light operating context
1. Detach the turbine to be set from the holder.
2. Press and hold the function key 2 / Setup (> 2 s).
Ä On the SIM display, the operating context Instrument light
appears. The selection focus Voltage (U) flashes.