Checking And Adjusting The Operating Pressures Of The Instruments; Factory Settings; Setting The Electric Motors - Dentsply Sirona Sinius Installation Instructions Manual

Hide thumbs Also See for Sinius:
Table of Contents

Advertisement

Dentsply Sirona
Installation Instructions Sinius / Sinius CS / Sinius TS
63 22 627 D3561
D3561.031.01.15.02 07.2018
5.2 Checking and adjusting the operating pressures of the instruments
10. Adjust the cuspidor flushing function:
Adjust control nozzle RD35 with the knurled knob until the cuspidor
bowl is rinsed evenly.
11. If the customer wishes, the factory-set suction power can be adjusted
with control nozzle RD33.
5.2
Checking and adjusting the operating pressures
of the instruments
5.2.1

Factory settings

A
Blowing air
B
Spray water
IMPORTANT
Factory setting
The operating pressures for the instruments are set at the factory.
Adjust the driving air pressure to the manufacturer's specifications for
the high-speed handpiece.
The driving air pressure can be set exactly using the four-way
measuring head (Order No. 36 66 336).
Factory settings for running high-speed handpieces (Sirona T1/T2/
T3/SIROPure)
Driving air on four-way measuring
head (TL)
5.2.2

Setting the electric motors

The cooling air pressure for the electric motor results from the driving air
setting of the turbine. A turbine must be fitted instead of the motor to
perform the setting.
5 Commissioning, demo mode and repacking
C
Spray air
D
Return air
2.7 bar ± 0.1 bar
129