4 Preparation
28
Instructions for Use Midwest Tradition / Tradition Pro / Quiet-Air
4.3

Before each patient

CAUTION!  Insufficient cooling leads to overheating of the
preparation site and to damage to the tooth substance. Ensure that
the water content is ≥ 50 ml/min.
1.
Purge the water and air channels for 30 seconds.
2.
Screw the high-speed handpiece onto the hose.
3.
Insert the bur.
4.
Set a sufficient amount of cooling water (> 50 ml/min).
5.
Use filtered water only (< 80  µm).
6.
Check the nozzles for blockages and lime deposits, for example,
and clean the nozzles if necessary.
67 27 155  D3673.201.10.02.02