Preparation; Initial Start-Up And Longer Breaks In Use; Prior To Starting The Work Day; Before Each Patient - Dentsply Sirona Midwest T1 Energo Instructions For Use Manual

Table of Contents

Advertisement

Available languages

Available languages

Dentsply Sirona
®
Instructions for use Midwest
T1/T2 Energo™
68 19 200 D3751
D3751.201.03.02.02    2023-11
4

Preparation

4.1

Initial start-up and longer breaks in use

➢ Clean and sterilize the instrument and accessories prior to startup 
[→ 18].
➢ Clean and maintain the instrument after longer breaks in use 
[→ 15].
4.2

Prior to starting the work day

CAUTION!  Risk of contamination! Purge the water and air
channels for 30 seconds.
4.3

Before each patient

1.
CAUTION!  Purge the water and air channels for 30 seconds.
2. NOTICE!  Stop the motor.
3.
CAUTION!  Do not attach the instrument until the motor has
stopped.  [→ 12]
4. Insert the preparation tool  [→ 13].
5.
CAUTION!  Set a sufficient amount of cooling water (> 50 mL/
min).
6.
CAUTION!   Use filtered water only (< 50 µm).
7.
CAUTION!   Check the nozzles for blockages and lime deposits,
for example, and clean the nozzles if necessary  [→ 17].
8. Check the fiber optic surface and clean it if needed  [→ 17].
4 Preparation
4.1 Initial start-up and longer breaks in use
11