5.1 Switching on the power

Before connecting to power, please check that the system has been correctly installed.
This includes:
• The device must be mechanically safe and mounted in compliance with the regulations.
• The power connections must have been made in compliance with the regulations.
• The electrical terminal compartments must be secured and the covers have been screwed
on.
• Check that the electrical operating data of the power supply are correct.
• Switching on the power.

5.2 General instructions for programming

After installation of the flow sensor(s) and the electrical connection of the converter, the device
can be switched on and is now ready to be programmed.
Start installation menu
Start installation menu
Start installation menu
Start installation menu
• Connect converter to power supply and power up converter.
Figure 5-1: First and second page appear intermittently
• Keep left button ">" pressed, until in display appears "release key now".
CAUTION!
If you program the diameter, use the outer diameter of the pipe.
For improved accuracy fill in as much details as possible.
Fill in the actual transducer distance at menu X7.2.6 (and X8.2.6 if applicable)
Run the optimization loop until the transducer distance changes no more than 0.5%.
INFORMATION!
For the description of the operating keys > ↓↑^ and other information refer to Display and
operating elements on page 101