VERSAFLOW SONIC 1000/TWS 9000

3.8.5 Mechanical installation of large version

INFORMATION!
You need a calculator, measuring band and pen & paper to install a large version.
The large sensor rail consists of an up and a down rail and has a transducer and a cable which
has to be connected directly to the converter or via a cable (splitter) box. A set of large sensor
rails can be installed as a V mode (reflected path), with both sensor rails on the same pipe side
or in a Z mode (direct path) with 1 sensor rail-mounted precisely on the opposite pipe side.

3.8.6 Installation of the UP sensor rail

CAUTION!
Make sure that you install the sensor rail centered to the pipe.
INFORMATION!
It is recommended to use and install extra fixing units and the cable (splitter) box, for pipe
diameters > DN2000 (as shown below).
Figure 3-14: Installing the large rail
1 Align the UP rail with the pipeline.
2 Optional fixing units for large diameters
3 Turn screws clockwise to secure.
4 Mark the position.
5 Optional cable (splitter) box
34-VF-25-07 iss.2 GLO Nov 11 US