OPERATION
6

6.3 Function tables

INFORMATION!
The following tables describe the functions of the standard device with HART
The functions for Modbus, are described in detail in the corresponding supplementary
instructions.
Depending on the device version, not all functions are available.

6.3.1 Menu A, Quick Setup

No.
Function
A1 Language
A1
language
A2 Tag
A2
Tag
A3 Reset
A3
reset
A3.1
reset errors
A3.2
totalizer 1
A3.3
totalizer 2
A3.4
totalizer 3
1
A4 Analog outputs (only for HART
A4
analog outputs
A4.1
measurement
A4.2
unit
A4.3
range
A4.4
low flow cutoff
A4.5
time constant
120
Setting / Description
language selection depends on the device version.
measuring point identifier (Tag no.) (also for HART
the LCD header (max. 8 digits).
reset errors? Select: no/yes
reset totaliser? Select: No / Yes
reset totaliser? Select: No / Yes
Reset Totaliser? Select: No / Yes
®
)
applicable to all current outputs (terminals A, B and C), frequency outputs
(terminals A, B and D), limit switch (terminals A, B, C, and / or D) and the 1st
display page / line 1
1) select measurement: Volume Flow / Velocity of Sound/ / Mass Flow /
Flow Speed / Gain / SNR / diagn flow speed, diagn VoS, diagn gain, diagn
SNR
2) use for all outputs? (also use this setting for Fct. A4.2...A4.5!)
Setting: no (applies only to the main current output ) / yes (applies to all
analogue outputs)
selection of the unit from a list, depending on the measurement
1) setting for main current output (range: 0...100%)
setting: 0...x.xx (format and unit, depending on measurement, see A4.1 and
A4.2 above)
2) Use for all outputs? Make setting, see Fct. A4.1 above!
1) setting for main current output (sets output value to "0")
setting: x.xxx ± x.xxx% (Range: 0.0...20%)
(1st value = switching point / 2nd value = hysteresis),
condition: 2nd value ≤ 1st value
2) Use for all outputs? Make setting, see Fct. A4.1 above!
1) setting for main current output (applicable to all flow measurements)
setting: xxx.x s (range: 000.1...100 s)
2) use for all outputs? Make setting, see Fct. A4.1 above!