Parameters For The Basic Configuration; Units - Honeywell VERSAFLOW SONIC 1000/TWS 9000 Handbook

Ultrasonic clamp-on flowmeter for liquids
Table of Contents

Advertisement

VERSAFLOW SONIC 1000/TWS 9000

9.5.2 Parameters for the basic configuration

There are parameters, such as measurement of counters, selection of diagnosis values and
setting of pressure and temperature correction, which require a warm start of the device
following data changes before other parameters may be written. Depending on the operating
mode of the host system (online/offline mode) these parameters need to be treated differently.
In online mode the settings should be changed with the corresponding online methods only, in
order to perform the warm start immediately and update dependent parameters automatically
afterwards.
In the menu tree these methods are located below the related parameters (e.g. in a counter
menu the method "Select Measurement" below the parameter "Measurement").
In offline mode the parameter "Online Mode?" in the "Detailed Setup / HART" menu should be
set to "no" before the configuration settings are changed as desired. Before writing the entire
offline configuration data set to the device the method "Prepare Parameter Download" in the
"Detailed Setup / HART" menu should be executed. This method writes the parameters for basic
configuration to the device and performs a warm start afterwards.
INFORMATION!
The Emerson Field Communicator and Simatic PDM execute this method automatically before
sending a configuration or performing a "Download to Device" respectively.

9.5.3 Units

There are separate physical units settings for configuration parameters and HART
dynamic/device variables. The units settings for configuration parameters are the same as on
the device's local display. They are available in the menu "Detailed Setup / Device / Units". For
each HART
in the menu "Detailed Setup / Process Input / HART". The different units settings can be aligned
with the method "Align HART Units" in the menu "Detailed Setup / Process Input / HART".
34-VF-25-07 iss.2 GLO Nov 11 US
®
dynamic/device variable the physical unit can be set individually. They are available
DESCRIPTION OF HART INTERFACE
www.process.honeywell.com
9
®
195