MCGV Fuse

Error ID: 2043
Valve Fuse 0: Channels A and B
Valve Fuse 1: Channels C and D
The gradient valve in the quaternary pump has drawn excessive current
causing the electronic fuse to open.
Probable cause
1
2
3

AIV Fuse

Error ID: 2044
The active-inlet valve in the module has drawn excessive current causing the
inlet-valve electronic fuse to open.
Probable cause
1
2
3
1260 Infinity QuatPump VL User Manual
Defective gradient valve.
Defective connection cable (front panel to
main board).
Defective main board.
Defective active inlet valve.
Defective connection cable (front panel to
main board).
Defective main board.
Error Information
Module Error Messages
Suggested actions
Restart the quaternary pump. If the error
message appears again, exchange the gradient
valve.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Restart the module. If the error message
appears again, exchange the active inlet valve.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.