2 Hardware Installation (Differences)
Installing the Column in the GC/MSD Interface
Installing the Column in the GC/MSD Interface
1 Carefully unwind the outlet end of the GC column until the 3 o'clock clip
is reached.
2 Slide an interface column nut (Part No. 05988-20066) and ferrule
(Part No. 5062-3508) onto the outlet end of the GC column.
The tapered end of the ferrule must point towards the nut.
3 Slide the column into the GC/MSD interface until the column protrudes
into the analyzer chamber at least 5 cm.
4 Adjust the length of the column from the 3 o'clock clip to the back of the
interface column nut to be 22 – 28 cm.
5 Hand tighten the interface nut.
6 Carefully close the oven door while observing to see that the column does
not develop sharp bends nor touch the oven walls/floor. Try this procedure
several times.
|Ophp_4dp4c4dazp
amFO"4Fm¤}
Oven open
34
¢¢J¢G4h
Oven closed