Download Print this page

Power Output Board - Leica CM3050 S Service Manual

Cryostat
Hide thumbs Also See for CM3050 S:

Advertisement

Chapter 3
General information:
The power output board controls the following de-
vices:
- Chamber illumination
- Heating - Screen
- Heating - Window
- Heating - Tray
- Heating - Pipe defrost
- Solenoid valve MV1 (German abbr. MV1)
Chamber
- Solenoid valve MV2 (German abbr. MV2)
Shut-off valve - specimen
- Solenoid valve MV3 (German abbr. MV3)
Control valve - specimen
- Fan control
- Heating cartridge rectifier
- Solid state relay compressor (German abbr.
'HBR')
Only the power output board for the 100V mains po-
wer input version has a different circuit. The diffe-
rence lies in the control electronics for the solid
state relay, which controls the compressors.
Follwing please find a description of the fuses for
the power output board and for the protected de-
vices.
Service Manual / Leica CM3050 S
Grundsätzliche Informationen:
Die Lastschaltkarte steuert folgende Komponenten
an :
- Beleuchtung Kammer
- HZ Blende
- HZ Scheibe
- HZ Wanne
- Rohrabtau-HZ
- Magnetventil MV1
Kammer
- Magnetventil MV2
Absperrventil Objekt
- Magnetventil MV3
Regelventil Objekt
- Lüftersteuerung
- Heizpatronengleichrichter
- Halbleiterrelais (HBR) Verdichter
Die Lastschaltkarte weist einen schaltungstechni-
schen Unterschied nur in der Version für 100V Netz-
spannung auf. Der Unterschied liegt in der Ansteu-
erelektronik für das Halbleiter-Relais der Verdichter.
Nachfolgend sind die Sicherungen der Lastschalt-
karte und die abgesicherten Verbraucher aufge-
führt.
Version 1.0 12/00
© Leica Microsystems

3.11 Power output board

Page 25

Advertisement