Cleaning The Pick&Place Module; Exchange Suction Cups - Leica HistoCore SPECTRA CV Instructions For Use Manual

Coverslipper
Table of Contents

Advertisement

7
Cleaning and Maintenance
7.2.10 Cleaning the Pick&Place module
• For handling the Pick&Place module, see
Clean the following areas of the Pick&Place module with a lint-free cloth soaked with xylene:
1. Check the front and rear suction cups
damages. Remove dirt with a lint-free cloth and a cloth soaked in xylene. Deformed or damaged
suction cups must be replaced
2. Carefully clean the skids
3. Control the coverglass sensor pin
L Dispose of glass shards, splinters and dust in accordance with laboratory specifications.
Note
At the bottom of the Pick&Place module is the coverglass sensor pin between both suction cups
(→ Fig. 91-3). The coverglass sensor pin can be moved freely if you can move it up and down by
touching it lightly with your fingertip.

7.2.11 Exchange suction cups

1. Unplug the deformed and/or dirty suction cups from the Pick&Place module and dispose in
accordance with laboratory regulations.
2. Remove new suction cups from the package.
3. Attach the suction cups at the front
module using a bent tweezers (→ Fig. 92-1).
114
(→ P. 86 – 6.3.5 Inspect Pick&Place
(→ Fig. 91-2)
(→ P. 114 – 7.2.11 Exchange suction
(→ Fig. 91-1)
on the left and right on the Pick&Place module from dirt.
(→ Fig. 91-3)
2
1
3
Fig. 91
(→ Fig. 92-2)
at the Pick&Place module for dirt and
cups).
for glued-on materials and mobility.
2
and rear
(→ Fig. 92-3)
module).
of the Pick&Place
Version 2.1, Revision K