Settings
64

11.1.5 Settings for D.O. sensors

The settings are available in the menu for measurement and calibration set-
tings. To open the settings, display the required measured parameter in the
measured value display and press the
tings, switch to the measured value display with
Default settings are printed in bold.
Menu item
Calibration /
Calibration record
Calibration /
Calibration data
storage / Display
Calibration /
Calibration data
storage / Output to
USB
Calibration /
Calibration interval
Calibration /
Comparison meas.
<OK>
Possible
Explanation
setting
Displays the calibration record of
-
the last calibration.
Displays the calibration record.
-
Further options:
 Scroll through the calibration
records with
 Output the displayed calibration
record to the interface with
<F2>
 Output all calibration records to
the interface with
output].
 Quit the display with
[Back] or
 Switch directly to the measured
value display with
Outputs the calibration records to
-
the interface.
Calibration interval for the D.O. sen-
1 ... 14 ...
sor (in days).
999 d
The meter reminds you to calibrate
regularly by the flashing sensor
symbol in the measuring screen.
Enables to adjust the measured
On
value with the aid of a comparison
Off
measurement, e.g. Winkler titration.
For details, see section 9.2 C
BRATION
MU 6100 H
key. After completing the set-
<MODE>
.
<
><
>
/[USB output].
<F2__>
<F1>
<OK>
.
<MODE>
, page 50.
ba77077e06
.
/[USB
/
.
-
ALI
05/2021