MU 6100 H
Application tips
ba77077e06
05/2021
The reference temperature and temperature compensation are set
in the Measurement menu for the parameter, conductivity (see sec-
tion 11.1.7 S
66).
Set the temperature compensation suitable for your test sample:
Test sample
Natural water
(ground water, surface
water, drinking water)
Ultrapure water
Other aqueous solutions
Salinity (seawater)

11.1.10 Setting the TDS factor

The factor to calculate the total dissolved solids is set to 1.00 in the delivery
condition.
You can adjust this factor to meet your requirements in the range 0.40 ... 1.00.
The factor is set in the menu for the parameter TDS (Measurement / TDS fac-
tor).
ETTINGS FOR CONDUCTIVITY MEASURING CELLS
Temperature compensation
nLF
according to EN 27 888
nLF
according to EN 27 888
Lin
Set linear temperature coeffi-
cient 0.000 ... 10.000 %/K
Automatic nLF according to
IOT (International Oceano-
graphic Tables)
Settings
, page
Display
nLF
nLF
Lin
Sal, nLF
69