Criteria for a stable
measured value
Freezes the mea-
sured value (HOLD
function)
ba77077e06
05/2021
page 71).
The Stability control function checks whether the measured values are stable
within the monitored time interval.
Measured parameter
D.O. concentration
D.O. saturation
D.O. partial pressure
Temperature
The minimum duration until a measured value is assessed as stable is the mon-
itored time interval. The actual duration is mostly longer.
With the HOLD function, you can freeze the current measured value. The dis-
played measured value stops changing until you switch the HOLD function off.
1.
Freeze the measured value with
The [HOLD] status indicator is displayed.
2.
Release the frozen measured value again with
The HOLD function is switched off.
The [HOLD] status display disappears.
9.1.2

Measuring the temperature

For reproducible D.O. measurements, it is essential to measure the tempera-
ture of the test sample.
The temperature is automatically measured by the temperature sensor
(NTC30 or Pt1000) integrated in the sensor.
Time interval
Stability in the time
interval
Δ : better than 0.05 mg/l
20 seconds
Δ : better than 0.6 %
20 seconds
Δ : Better than 1.2 mbar
20 seconds
Δ : better than 0.5 °C
15 seconds
<HOLD>
.
Dissolved oxygen
<HOLD>
.
49