Manual Calibration (Anycal) - wtw VWR MU 6100 H Operating Manual

Table of Contents

Advertisement

pH value
32
21.
Wait for the measurement with stability control to be completed or termi-
nate the stability control and take over the calibration value with
The calibration display for the next buffer appears (voltage display).
22.
If necessary, use
Continue calibrating using the next buffer with
Calibration is automatically completed after the last buffer of a buf-
fer set has been measured. Then the calibration record is dis-
played.
The calibration line is determined by linear regression.
6.2.4

Manual calibration (AnyCal)

Make sure that in the sensor menu, Buffer menu, the AnyCal buffer set is
correctly selected (see section 11.1.1 S
60).
Use any one to five buffer solutions in ascending or descending order.
If single-point calibration was set in the menu, the calibration proce-
dure is automatically finished with the measurement of buffer solu-
tion 1 and the calibration record is displayed.
1.
Connect the pH electrode to the meter.
The pH measuring window is displayed.
2.
When measuring without temperature sensor:
– Temper the buffer solution, or measure the current temperature.
– Enter the temperature value in the menu.
3.
In the measured value display, select the measured parameter pH or mV
<MODE>
with
.
4.
Start the calibration with
The calibration display appears (voltage display).
pH
AnyCal
15.02.2014
5.
Thoroughly rinse the electrode with deionized water.
6.
Immerse the electrode in the first buffer solution.
<MODE>
to finish calibration or
ETTINGS FOR P
<CAL>
.
1
Buffer
0.0
mV
24.8
°C
14:15
MU 6100 H
<OK>
<OK>
.
H
, page
MEASUREMENTS
ba77077e06
05/2021
.