Data Memory; Manual Storage - wtw VWR MU 6100 H Operating Manual

Table of Contents

Advertisement

Data memory

74
12
Data memory
You can transmit measured values (datasets) to the data memory:
 Manual data memory (see section 12.1 M
 Automatic storage at intervals (see section 12.2 A
, page 75)
AT INTERVALS
With each data storage process, the current datasets of the sen-
sors indicated on the display are transmitted to the interface at the
same time.

12.1 Manual storage

You can transmit a measurement dataset to the data memory as follows.
With each data storage process, the current datasets of the sensors indicated
on the display are transmitted to the interface at the same time.
<STR>
1.
Press the
The menu for manual data storage appears.
Manual data storage 4 von 200
15.02.2014 11:24:16
pH 7.0
24.8 °C AR +++
O2 7.80 mg/l
24.8 °C AR +++
ID number:
Continue
15.02.2014
Back
2.
If necessary, change and confirm the ID number (1 ... 10000) with
<
><
>
<OK>
and
The dataset is stored. The meter switches to the measured value dis-
play.
key shortly.
1
14:15
.
, page 74)
ANUAL STORAGE
UTOMATIC DATA STORAGE
ba77077e06
MU 6100 H
05/2021