Dissolved oxygen
52
4.
Start the measurement with
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
5.
Wait for the end of the measurement with stability control or accept the
calibration value with
The factor that was set last is displayed.
O
2
7.9
OxiComp
15.02.2014
<
> <
6.
Using
centration value to the nominal value (value of the comparison mea-
surement). Subsequently, accept the correction factor with
The meter switches to the measured value display.
The status display [Factor] is active.
<OK>
.
<OK>
.
mg/l
0.973
08:00
>
, set the correction factor to adjust the displayed con-
MU 6100 H
<OK>
.
ba77077e06
05/2021